Accurate Balance

Accurate Balance MA-50/110 Series: Touch Screen Moisture Analyzer

Download   

             00961 1 894 243

Description

Specifications:

Model MA-50-1/110-1 MA-50-5/110-5     MA-50-10/110-10
Capacity (g) 50g/110g 50g/110g 50g/110g
Readability (g) 0.001g 0.005g 0.01g
Repeatability 0.02-0.1% (>=3g)
Suggested sample 3-10g
Heating time 0-99min
Environment 10-40°C
Interface RS232
Heating Temp. 50°C to 180°C
Moisture 0-100%
Pan 100mm
Outline 330mm*180mm*140mm
N./G. Weight 3Kg/4.5Kg
Heating Halogen
Packing 47*31*33cm/carton

Standard Attachments:

  • Machine
  • Bracket
  • Power cord
  • Weighing plate
  • Calibration weights 50g
  • Tweezers
  • Operation manual

How to use:

  • Moisture meter is precision analysis instruments, users should be careful operation, according to the sample shDUld be evenly spread in the pan
  • The typical sample is 3-5g, tare before weighing
  • Please set the heating parameters (heating temperature, end), then weight the sample
  • In order to ensure the aCcuracy of the test, the sample is the best powder state, large partiCles of the sample should be used to crush the
  • In order to measure correctly, the second use to wait for moisture meter cooling 50”ñ

Turn on/Off:

  • In the off state, touch screen, start moisture meter, the instrument inside the self inspection instrument after the end of the end of the main interface

Operation:

Calibration of weighing: In order to improve and ensure the accuracy of the weighing data, weight calibration should be carried out before the first weighing or the accuracy of the requirement.

How to calibrate:

  1. Clean the sample tray, the sample lray should place well. fouch the weight calibration key in the main
  2. Touch the calibration key and start the weight
  3. PlaCe the weights when the display show 50g 0r 1D0g, then close the cover.

 