Description
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.
Features for TopPette
* Light weight, ergonomic, low force design
* Digital display clearly reads volume setting
* The pipettes cover volume range of 0.1μl to 10ml
* Easy to calibrate and maintain
* The lower part is autoclavable
* Design helps avoid repetitive strain injuries
Features for Multi-channel pipette
* 8 and 12 channel pipettes are available for standard 96-well plate
* Dispensing head rotates ensure optimum pipetting convenience
* Individual piston/tip cone assemblies allow easy repair and maintenance
* Compound material tip cone design allows visual seal vertification
* Universal pipette tips can be used
Technical Parameters:
8-channel Adjustable Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume
(μl) |
Maximume Permissible Systematic error
(Inaccuracy) |
Maximume Permissible Random error
(Inaccuracy) |
||||
% | μl | % | μl | |||||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 | ||
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 | ||
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 | ||
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||||
50 | 1.50% | 0.75 | 0.80% | 0.4 |