Specifications |
Resistance |
Nominal 100 Ω (± 0.1 Ω) |
Temperature Coefficient |
0.003925 Ω/Ω/°C nominal |
Temperature Range |
–100 °C to 156 °C |
Transition Temperature |
–100 °C to 156 °C |
Drift Rate |
± 0.01 °C per year maximum at 0 °C, when used periodically at max temperature |
Sheath Material |
Inconel™ 600 |
Leads |
PTFE-insulated, silver-plated stranded copper, 22 AWG. |
Termination |
Specify. See Models and Accessories. |
Calibration |
Includes manufacturer’s NVLAP-accredited, lab code 200706-0, calibration and table with R vs. T values in 1 °C increments from –80 °C to 156 °C. ITS-90 coefficients included. |
Calibrated Accuracy†(k=2) |
± 0.05 °C over the full range |
Cable Length |
6.1 meters (20 ft) |
Size |
6.35 mm (0.25 in) dia. x 152 mm (6 in) |