FLUKE

Fluke Biomedical MaxO2 PLUS AE: Oxygen Analyzer

Download   

             00961 1 894 243

Category:

Description

Key features:
• One-touch calibration, with reminder
• Long battery life (approx. 5,000 hrs)
• Impact resistant and drip proof
• External MAX-250E Oxygen Sensor

The MAXO2+AE is an oxygen analyzer that measures the oxygen concentration in a flow of gas from a
medical gas source or through a medical gas-flow device such as a ventilator or anesthesia system, or
within an infant incubator. It is handheld and rugged to suit the needs of portable use. The MAXO2
+ AE comes equipped with a two-year warranty on both analyzer and sensor.

Measurement range 0 % to100 %
Resolution 0.1 %
Accuracy and linearity 1 % of full scale at constant temperature, RH and pressure when calibrated at full scale
Total accuracy ± 3 % actual oxygen level over full operating range
Response time 90 % of final value in approx. 15 sec at 23 °C
Warm-up time None required
Power supply
Battery life Approx. 5000 hours with continuous use
Low battery indication “BAT” icon displayed on LCD
Sensor type Maxtec® MAX-250E for AE model
Expected sensor life >900,000 O2 % hours minimum, two years in typical medical applications
Power requirements Two, AA Alkaline batteries
Environmental requirements
Operating temperature 15 °C to 40 °C (59 °F to 104 °F)
Storage temperature -15 °C to 50 °C (5 °F to 122 °F)
Atmospheric pressure -800 mBar to 1013 mBar
Relative humidity Operating range: 0 % to 95 % (non-condensing)
General information
Dimensions (LxWxH) 38 mm x 76 mm x 914 mm (1.5 in x 3 in x 36 in)
Weight 285 g (0.6 lb)

Additional information

Brands
Labels