Fluke Calibration

Fluke Calibration P3000 Series: Pneumatic Deadweight Testers

Fluke

Download         Download

             00961 1 894 243

Description

Features:
• Pressure ranges to 20,000 psi (1,400 bar)
• Accuracy better than 0.015% of reading (Increased accuracy option of 0.008%)
• Mineral oil and distilled water models
• Dual piston models allow calibration over a wide range
• Psi, bar, kgf/cm2, kPa and MPa ranges available
• Dual piston models for calibration over a wide pressure range
• Built-in lever style hand pump now standard on all hydraulic models
• Mounted spirit level and adjustable feet
• High quality screw press for fine pressure control
• Test station design with O-rings eliminates the need for PTFE tape or wrenches
• Acrylic reservoir provides visibility of fluid level and quality
• Improved case with springloaded latches for hood
• Built-in drain plug to remove old fluid
• Sturdy weight box with hinged lid and side handles for easy transport

The P3000 Series is the culmination of over 50 years experience in the production and design of primary pressure standards. With features designed to improve accuracy and performance, increase reliability and simplify operation these deadweight testers can be used to calibrate virtually any pressure sensing device, including transducers, transmitters, gauges or pressure switches.
Each instrument is supplied with a detachable lid which makes it neat, compact and easily portable.
The weight masses are stored in a  high-quality case with a self locking mechanism to protect them during transit.
All units are provided with an accredited calibration report, weight mass details, M14 x 1.5, M20 X 1.5, 1/8 inch, 1/4 inch, 3/8 inch and 1/2 inch NPT and BSP female adaptors, operating fluid (where applicable) and spare seals.
The piston/cylinder assemblies are manufactured to the very highest standards and are provided standard with ISO/IEC 17025 accredited calibration reports.

Specifications:

Accuracy
Accuracy1 ±0.015% of reading (±0.008% optional)
1. Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range.
Mass and PCU Materials of Construction
Standard weight material Series 3 non-magnetic austenitic stainless steel
Weight density 7.8 g/cm³
Optional fractional weights Solution heat treated aluminum
Weight density 2.7 g/cm³
Models P3011, P3012, P3013, P3022, P3023, P3025 (V) Piston material Ceramic
Cylinder material Martensitic steel
Coefficient of expansion 11 ppm/°C
Models P3014, P3015, P3025 (P) P3031, P3032 (P & V) Piston material Tungsten carbide
Cylinder material Martensitic steel
Coefficient of expansion 16.5 ppm/°C
Model P3016 Piston material Tungsten carbide
Cylinder Material Tungsten carbide
Coefficient of Expansion 11 ppm/°C
General
Test port adaptors 1/8, 1/4, 3/8 and ½ NPT or BSP
Instrument weight 24 lbs
Instrument size (W x D x H) 17.5 x 12 x 8.5
Mass set weight (typical) 65 lbs
Minimum Standard Weight Increments
P3012, P3022 Pressure 1 inH2O, (5 mbar)
P3013, P3023 Pressure 2 inH2O, (10 mbar)
P3014, P3015, P3025 Pressure 1 psi, (0.1 bar)
P3016, P3031, P3032 1 psi, (0.1 bar)
P3011, P3022, P3023, P3025 Vacuum 0.2 inHg (10 mbar)
Optional Fractional Weights
P3014, P3015, P3025 Pressure 0.1 psi (0.01 bar)
Built-in Hand Pumps
Pressure mode Max pressure 300 psi (20 bar)
Vacuum mode To 90% vacuum
Operating Fluid for Liquid Lubricated Models
55-655 Mineral oil (standard)
Krytox For oxygen-safe applications (optional)
Options
PressCal Software Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments.
Krytox As operating fluid for liquid lubricated gas operated models.

Additional information

Brands

Labels